YMC Capillary Column
YMC Capillary columns are compatible with Nano-LC/Micro-LC systems. Miniaturisation of liquid chromatography in combination with mass spectrometry offers improvements in sensitivity, especially at low concentration levels and dramatically reduced solvent consumption, compared to conventional HPLC or UHPLC. They are available either with 1/16" connections (10-32 thread) or with 1/32"connections (6-40 thread).
To meet the requirements of MicroLC/CapillaryLC/NanoLC YMC offers capillary columns specifically designed to use with the corresponding chromatography systems.
Suitable for high-sensitive analysis
Great selection of functional groups
Operating pressure up to 60 MPa (1.9 µm) / 55 MPa (2/3/5 µm)
Guard/trapping columns available (supplied with column coupler)
Parameter
|
Specification
|
||||
IDs* | 75, 100, 300, 500 µm | ||||
Lengths | 50, 75, 100, 150 mm | ||||
Particle sizes | 1.9 - 5 µm | ||||
Phases
|
all fully porous YMC phases, e.g.
|
||||
Pressure limits**
|
1.9 µm: 600 bar / 8,700 psi
2-5 µm: 550 bar / 7,975 psi
|
||||
Connections
|
1/16" (10-32 thread)
1/32" (6-40 thread)
|
||||
Guard/Trapping columns
|
5 x 0.3 mm ID
or 5 x 0.5 mm ID
|
*0.075 / 0.1 mm ID for particle sizes >2 µm. ** Hardware pressure limited to 690 bar / 10,000 psi.
Guard/Trapping columns
Guard columns can be connected directly to the analytical column using the column couplers supplied.
Column coupler
A column coupler is supplied with every pack of capillary guard cartridges to guarantee the optimum connection with low dead volume. A polymer-based (PCTFE) coupler is provided for 1/16” columns (bottom), while a stainless steel coupler is provided for 1/32” columns (top). Every coupler can be purchased separately if required.
Title | Document Type | Language | Phase Type | Download |
MicroLC detection of gluten in flour and cookies with YMC-Triart C18 | Application Note | E | Download | |
MicroLC detection of veterinary drug residues in food with YMC-Triart C18 | Application Note | E | Download | |
YMC Capillary Columns Column Care and Use | Care and Use Instructions | E | Download | |
Analytical stationary phases for HPLC and UHPLC from YMC | Overview | E | Download |